OmniFlux Complete Guide

Video tutorial: watch how to set up and use OmniFlux

Part 1: The device

OmniFlux measures roast color in real time so you can Roast by Color. Mounted on its stand outside the roaster’s viewing window, it records a color curve through the whole roast, marks yellowing, first crack and second crack, and alerts you when the beans reach the target color you set. Off the stand, with the Color Test Adapter fitted, it measures the average roast color of roasted beans or grounds on the sample tray.

A color curve carries over from one roaster to another in a way a temperature curve does not, because every machine measures temperature slightly differently. That makes it possible to match a sample roast on a production roaster by color.

OmniFlux on its stand, tracking a roast through the viewing window.
OmniFlux on its stand, tracking a roast through the viewing window.

What’s in the box

omniflux_box

OmniFlux
Color Test Adapter (metal)
Sample tray
Two-color calibration card
Cleaning brush and scraper
Cleaning cloth
Stand
Screwdriver
USB-C charging cable
Toolbox
User manual with warranty card, and QR code card

Sold separately: the USB light and the PT100 temperature probes (two ports, T1 and T2). There is no power adapter in the box. Use a USB adapter rated 5 V / 0.8 A.

Know your OmniFlux

The back, bottom and front of OmniFlux.
The back, bottom and front of OmniFlux.

Display screen - the touchscreen on the front. Every mode, setting and result is on it.

Power button - the only physical button. It powers OmniFlux on and off, and starts or stops the current test.

Optical module - the lens on the back. The Color Test Adapter fits over it for static color tests.

Infrared laser - the red dot you aim at the beans. Never look into it.

Distance measurement sensor - hidden on the back. It measures the distance to the target, and tapping the READY screen enters Color Test or Roast Track depending on that distance.

Infrared temperature sensor - on the back. It measures bean temperature during Cool Track, and spot temperature in the infrared temperature utility.

Temperature and humidity sensor - on the bottom. It measures the temperature and humidity of the room, not the beans.

TEMP1 and TEMP2 - the two temperature probe ports on the bottom, for PT100 probes.

Bracket connection point - the threaded hole on the bottom where the stand’s quick-release head attaches.

USB-C port - charging and data transfer, on the bottom.

USB power output - a 5 V port that powers the USB light.

The bottom of OmniFlux: TEMP1 and TEMP2 probe ports, the bracket connection point, and USB-C.
The bottom of OmniFlux: TEMP1 and TEMP2 probe ports, the bracket connection point, and USB-C.

Color Test, Roast Track and Cool Track

Tap anywhere on the READY screen to enter Color Test or Roast Track. Which one you get depends on the distance between OmniFlux and whatever is in front of the lens.

ModeWhat it measuresRange and accuracy
Color TestStatic test of roasted beans or grounds on the sample tray, with the Color Test Adapter fitted. Starts at about 7 cm.±0.1 Agtron, 0 to 150
Roast TrackLive color curve of the beans in the drum, through the viewing window. Starts at over 10 cm. Optional probes add bean and chamber temperature.±1 Agtron, 0 to 200, at 20 to 60 cm
Cool TrackInfrared temperature of the beans in the cooling tray after the drop, plotted as a curve.±1 °C, 0 to 320 °C, at about 20 to 25 cm
Color Test: fit the adapter, place OmniFlux on the sample tray, tap the READY screen.
Color Test: fit the adapter, place OmniFlux on the sample tray, tap the READY screen.
Roast Track: mount OmniFlux on the stand, tap the READY screen.
Roast Track: mount OmniFlux on the stand, tap the READY screen.

The button and the screen

ButtonWhat it does
Hold 3 seconds (off)Powers on. The first time, and after a factory reset, this is the only way to power on. After you choose a language, a short press works too.
Hold 3 seconds (on)Shows “Shut Down”. Release to power off.
Hold 10 secondsForces OmniFlux off if it does not shut down normally.
Short pressOn a test or calibration page, starts the current function, and a second press stops it. Anywhere else, returns to the READY screen.
omniflux_ready
On screenWhat it is
1Connection to CoffeeOS: a phone icon shows when CoffeeOS is connected.
2Battery level.
3Swipe from left to right to open the menu: calibration, settings and more.
4Swipe from right to left to open History, every previous measurement.
5The utility icon: distance, infrared temperature and the room’s environment.
6Tap anywhere to enter Color Test or Roast Track.

Settings

Swipe from left to right on the READY screen and tap Settings.

Settings, Device Info, Device Settings and System Settings.
Settings, Device Info, Device Settings and System Settings.

Device Info - the serial number, firmware version, and a QR code. Scan the QR code with your phone camera to download CoffeeOS, or scan it from inside CoffeeOS to connect.

Connections - under Device Settings. Bluetooth (on to connect to CoffeeOS), AirWave connection, alert volume, and the USB light switch.

Device Settings - also holds screen brightness, auto shut-off time, Flip Display (turns the whole screen 180° when OmniFlux is mounted upside down), date and time format, language (English, Japanese or Simplified Chinese), and factory reset. Hold Factory Reset until the bar finishes to clear every setting and all history.

System Settings - your Roast Track roaster profiles, the color standard (SCA or COMMON), and temperature in °C or °F.

Power and charging

The battery lasts about 4 hours and takes about 4 hours to charge through the USB-C port on the bottom. OmniFlux works while it charges. Start a session with the battery above 15%.

If OmniFlux will sit unused for a while, charge it fully first and top it up from time to time. Storing it with a flat battery can damage it.

Keep it clean

  • Wipe the lens with a soft, lint-free cloth every day before you measure. Fingerprints, dust and oil affect the reading.
  • Wipe the inside of the roaster’s viewing window regularly. Oil builds up on the glass and affects Roast Track.
  • Clean the sample tray with the brush or with water, dry it, and make sure no silverskin is left in it.

Calibrate the color reading

Color calibration standardizes OmniFlux against the two references on the calibration card. Do it before you use OmniFlux for the first time, then once a week.

1. Lay out the card
Take the two-color calibration card out of its packaging and lay it flat. Keep the two detection areas free of scratches.

omniflux_cal_card

2. Open calibration
Swipe from left to right on the READY screen, then tap Calibration, then Color Test Calibration.

3. Enter the card’s values
Enter the two values printed on your card and tap Next.

omniflux_cal_values

4. Fit the Color Test Adapter
Fit the adapter over the optical module.

omniflux_cal_adapter

5. Measure both references
Place OmniFlux on reference ① and tap Test ①. Move it to reference ② and tap Test ②. When both show a check mark, tap Calibrate.

omniflux_cal_measure

Replacing the card or the adapter: replacing the calibration card or the Color Test Adapter means sending OmniFlux back to DiFluid for recalibration.

Set up the stand

Clamp the stand
1. Clamp the stand
Clamp the stand to the edge of a table or another firm surface and tighten the knob.
Fit the connector
2. Fit the connector
Take the nut off the connector, place the nut on the stand, then press the connector firmly into the top of the stand.
Tighten both points
3. Tighten both points
Tighten the two points marked by the arrows.
Attach the quick-release head
4. Attach the quick-release head
Line up the positioning post on the quick-release head with the hole on the bottom of OmniFlux and tighten it with the supplied screwdriver.
Mount OmniFlux
5. Mount OmniFlux
Hold the button on the quick-release head and slide OmniFlux on. To take it off, hold the same button and lift gently.
OmniFlux mounted on the stand.
OmniFlux mounted on the stand.

Check that every knob on the stand is tight before you roast. A loose knob lets the angle shift, and the color curve shifts with it. Never shake the stand when you lift OmniFlux off.

Configure your roaster

Before the first roast with Roast Track, create a configuration for your roaster. During one roast, OmniFlux records the color it measures through the viewing window for the green beans, and for the roasted beans just before the drop. You then measure the cooled beans with the Color Test Adapter, and that measurement is used to correct the color measured through the glass on every later roast.

  • Most roasters need one configuration.
  • If you roast very different coffees, such as blends with a wide range of color, or different processes, create a separate configuration for each and name them so you can tell them apart (Roest A, Roest B).
  • Each roaster needs its own configuration.
  • “Default” is the factory configuration. You can pick it in the Roast Track pull-down settings, roast, and save it under your roaster’s name afterward.

1. Start a new configuration
Swipe from left to right on the READY screen, tap Calibration, then Roast Track Calibration, then New Roaster Profile. Enter your roaster’s name.

2. Set the stand
Adjust the stand’s height and angle for your viewing window and tighten every knob. Mount OmniFlux on the stand.

3. Aim the laser
Aim the red dot level or slightly downward at the center of the viewing window, where the bean pile is densest and moves least. That gives OmniFlux the largest area of beans to read.

4. Charge the green beans
Preheat the roaster fully. Charge enough green beans to cover the viewing window completely, then tap Next.

5. Record the green beans
The raw color value jumps at first. When it settles, after a few seconds, tap Record Green Beans. Then roast normally without touching OmniFlux.

omniflux_config_green

6. Record the roasted beans
Just before you drop, tap Record Roasted Beans, then drop immediately. Any delay between the two shifts the final color OmniFlux records.

7. Measure the cooled beans
Lift OmniFlux off the stand without moving the stand. When the beans have cooled, fit the Color Test Adapter, tap the Roasted Bean Color Value field, and measure the beans on the sample tray (see Prepare a sample). Enter the result, tap Done, then tap Build Profile.

What to watch for: the raw color and distance values jumping around during configuration is normal. The green beans’ raw value should be clearly higher than the roasted beans’ raw value. If the two are close, or only about 10 apart, there usually were not enough beans in the drum to cover the window.

After configuring: do not move the stand. Roast Track needs the same angle and distance on every roast as during configuration. With a small viewing window, leave the stand fixed in place so you never have to find the angle again.

Prepare a sample

How you fill the tray changes the color reading as much as anything else. For a quick reading, fill the sample tray and level it with the scraper. Use the deep side for whole beans and the shallow side for grounds.

Fill the tray, then level it flush with the scraper.
Fill the tray, then level it flush with the scraper.

For the most accurate reading:

1. Start with a clean tray
Brush or rinse the tray and dry it. No silverskin or debris should be left.

2. Fill and level
Fill the tray completely and scrape the surface flat.

3. Remove the outliers
Pick out defects and beans with an obvious color difference, and replace them with normal beans.

4. Flip silverskin down
On light and medium roasts where silverskin is clearly visible, turn each bean so the silverskin faces down. Level again so the beans are flush with the edge of the tray.

5. Repeat
Empty the tray, reload it and test again, 2 to 3 times. When the results are close, use the final value. If they still vary, reload a few more times and average them. Blends and coffees with a wide range of color need more loads.

Measure roast color

A Color Test on whole beans.
A Color Test on whole beans.

1. Fit the Color Test Adapter
Fit the adapter over the optical module.

omniflux_ct_adapter

2. Prepare the sample
Fill and level the sample tray (see Prepare a sample).

3. Place OmniFlux on the tray
Line the adapter up with the edge of the metal tray and set OmniFlux on it.

omniflux_ct_attach

4. Open Color Test
Tap the READY screen and choose whole beans or grounds as the sample type.

5. Tap New Test
New Test starts a new series and shows only this result.

omniflux_ct_newtest

6. Tap Add Test for more loads
Reload the tray and tap Add Test. The result shown is the running average of every test since the last New Test.

omniflux_ct_addtest
SituationTap
You switched to a different coffee, or from beans to groundsNew Test
You finished a series and are starting the next batchNew Test
You are measuring one batch in several portions to average themAdd Test
You are averaging one sample from several angles, by turning the tray or redistributing the groundsAdd Test

Whole bean and grounds: grounds usually read lighter than whole beans from the same batch. Measure both, and CoffeeOS shows the difference as Roast Delta (see Part 2).

Track a roast

omniflux_photo_roaster2

Before you charge. Mount OmniFlux on the stand at the same angle and distance as when you configured the roaster. Tap the READY screen to open Roast Track, then pull down from the top of the screen to open the pull-down settings.

omniflux_rt_ready
Roast Track pull-down settings.
Roast Track pull-down settings.

Roaster Profile - the roaster configuration for this roast.

Auto Start/Stop - on: tracking starts automatically when you charge, continues after the target color, and the result saves automatically when you charge the next batch. Suits back-to-back batches. Off: tap the start icon in the lower right after charging, and tracking stops when the target color is reached.

Color Target Alerts - sounds an alert when the beans reach your target color. Leave it on.

Temperature Probe T1 and T2 - turn on when a probe is connected. Display Curve switches the chart between color (Ag) and probe temperature (T1, T2).

USB Light - Off, On or Auto, when the USB light is connected.

Set the target color by holding the + and − buttons. Preheat the roaster fully, then charge. With Auto Start/Stop off, tap the start icon in the lower right as you charge.

Set the target color with + and −.
Set the target color with + and −.

During the roast. OmniFlux marks yellowing, first crack and second crack on the curve automatically, and you can add marks by hand. The screen shows the color value and crack count at each mark. You can change the target color at any time.

The color curve with yellowing, first crack and second crack marked.
The color curve with yellowing, first crack and second crack marked.

Before the drop. Use the color trend to see the drop coming. Note the time or color at the moment you drop, so you can find the drop point when you check the result.

At the drop. When the alert sounds, drop immediately. You don’t need to touch OmniFlux. Tap End Roasting in the pop-up afterward.

Cool Track. After the drop, tap the Cool Track icon in the lower right. Lift OmniFlux off the stand without moving the stand, hold it about 20 cm above the cooling tray, and aim the red dot at the beans. The bean temperature plots as a curve. Tap the flame icon to go back to the roast result.

If a roast doesn’t save: a record saves only when a full charge-to-drop cycle is detected. A hand blocking the laser at the drop, the laser pointing at the drop chute handle, or a deep drum that returns no signal can all stop the drop from being detected.

Correct the drop color

Compare the color OmniFlux tracked at the drop with the color of the cooled beans measured on the sample tray, and save the difference to the roaster profile. Done carefully, the deviation on later roasts stays within ±1 Agtron. Coffees that are uneven in color can deviate more.

1. Check the drop point
On the results page, confirm the time or color shown matches what you noted at the drop. If it does, go to step 2. If not, tap the results page to open trimming, and adjust the drop point in one of three ways: drag the dashed line, enter the drop color, or enter the drop time. Then tap the trim icon to cut the curve after the drop.

Tap the results page to open trimming and calibration.
Tap the results page to open trimming and calibration.

2. Measure the cooled beans
Prepare the cooled beans on the sample tray and fit the Color Test Adapter.

3. Recalibrate
Tap the drop color, choose the roaster’s name, tap Recalibrate, tap the final roast color field, and test. Reload and test 2 to 3 times, then tap Save.

omniflux_rt_recal

If OmniFlux tracked 75 at the drop and the cooled beans measure 78, the deviation is 3. The drop color updates to 78, and the next roast on that profile shows a more accurate color. If the beans measure 75, there was no deviation and nothing changes.

Pair with AirWave

Paired with OmniFlux, AirWave adjusts its fan speed automatically to the stage of the roast.

1. Power on both
Turn on AirWave and OmniFlux.

2. Set AirWave to Auto Mode
On AirWave’s home screen, hold the knob to open settings. Turn the knob right to Mode and press to select Auto Mode. Keep turning to Wireless Connection and press to turn it on.

3. Connect on OmniFlux
On OmniFlux, swipe from left to right on the READY screen, then tap Settings, Device Settings, Connection Settings, AirWave Connection. Tap AirWave’s name when it appears, then swipe right to exit.

4. Return AirWave to its home screen
Hold the knob on AirWave to exit settings.

5. Roast once by hand
Open Roast Track on OmniFlux and roast. Adjust AirWave’s fan speed by hand as the smoke changes, for example 70% at yellowing and 80% at first crack. OmniFlux records each change.

On every roast after that, AirWave repeats the fan speeds from that first roast. Turn the knob during any roast to change them, and the newest settings are the ones repeated.

Built-in utilities

Tap the utility icon in the top right of the READY screen.

omniflux_tools
UtilityWhat it measures
DistanceLaser distance, 5 cm to 50 m, ±3 cm. Measures the distance between OmniFlux and the roaster window when you set up the stand.
Infrared temperatureSpot temperature where the red dot points, −70 to 380 °C, ±1 °C. Use it on the roaster’s surface or the cooling tray.
EnvironmentAir temperature, humidity, pressure and altitude of the room you roast in.

With a PT100 probe connected, OmniFlux also measures probe temperature. Turn the probe on in the Roast Track pull-down settings to see it on the roast curve.

Roast color reference

AgtronCOMMONSCA
0 to 30EspressoVery Dark
30 to 40FrenchDark
40 to 50Full CityMedium Dark
50 to 60CityMedium
60 to 70DarkMedium Light
70 to 80MediumLight
80 to 90CinnamonVery Light
90 to 150LightExtremely Light

Higher Agtron values are lighter roasts. Set which standard OmniFlux displays in System Settings.

When something looks wrong

ProblemWhat to do
Won’t power on the first timeHold the button for 3 seconds. A short press only works after you choose a language.
Won’t connect to CoffeeOSTurn on Bluetooth in Settings, Device Settings, Connection Settings. On your phone, allow Bluetooth, location and every CoffeeOS permission. Restart OmniFlux and CoffeeOS.
Roast Track color is offCheck the red dot is on the beans, not the drum wall or the space above them. Work at about 40 cm from the viewing window: closer than that, heat from the roaster can make the reading unstable. Clean the window, check the stand hasn’t moved, and redo the roaster configuration if needed.
Color Test is offCalibrate with the card, then check sample prep: a level, full tray with defects and loose silverskin removed.
A roast didn’t saveSee “If a roast doesn’t save” under Track a roast.
Double-glazed viewing windowColor changes are hard to see through double glazing, especially with the roaster’s lamp on. Contact DiFluid support before relying on Roast Track.

For anything else, email service@digitizefluid.com with a description, a video of the problem, and the serial number. The serial number is on the right-hand side of OmniFlux.

Specifications

ItemValue
Color Test±0.1 Agtron, 0 to 150 Agtron, whole beans and grounds
Roast Track±1 Agtron, 0 to 200 Agtron, monitoring distance 20 to 60 cm, any angle with a clear optical path
Cool Track±1 °C, 0 to 320 °C, at about 20 to 25 cm
Infrared temperature±1 °C, −70 to 380 °C
Laser distance5 cm to 50 m, ±3 cm
BatteryAbout 4 hours, works while charging
ChargingAbout 4 hours, USB-C, 5 V / 0.8 A adapter (not included)
USB power output5 V
ConnectivityBluetooth to CoffeeOS and AirWave
LanguagesEnglish, Japanese, Simplified Chinese
WarrantyOne year from the date of purchase

Part 2: OmniFlux in CoffeeOS

CoffeeOS is free, with no purchase necessary. Connect OmniFlux to keep its firmware up to date, and to bring Color Test readings into the Roast Color Analyzer in CoffeeOS, where you can compare whole bean and grounds and link the result to the coffee in Bean Manager.

Connect OmniFlux

Download CoffeeOS here, or scan the QR code on the Device Info screen with your phone camera.

1. Turn on Bluetooth
On OmniFlux, open Settings, Device Settings, Connection Settings and turn Bluetooth on.

2. Open the Tools tab
Open the Tools tab in CoffeeOS with OmniFlux on and nearby. A pop-up offers to connect it.

3. Or scan the device
If no pop-up appears, open the quick menu in the top right of the Tools tab, tap Scan, and scan the QR code on OmniFlux’s Device Info screen.

Update the firmware

Connect to CoffeeOS regularly to check for new firmware. Updates add features and keep OmniFlux running smoothly. With OmniFlux connected, go to Me, Device Center, and tap OmniFlux.

Roast color in the Roast Color Analyzer

The Roast Color Analyzer shows whole bean color at the top and grounds color at the bottom. Measure both from the same batch in one session, and it calculates Roast Delta, the difference between the two.

1. Open a session
Open the Roast Color Analyzer and choose OmniFlux in the device selector at the bottom left of the Tool Controller.

2. Test whole beans
Fit the Color Test Adapter and prepare the whole beans. Tap Add Test and choose whole bean as the sample type.

3. Test grounds
Grind part of the same batch, prepare it in the shallow side of the tray, tap Add Test and choose grounds.

Grounds usually read lighter than whole beans, because the outside of the bean is darker than the inside. Roasters disagree on what Roast Delta should be, so the number to aim for is your choice.

For more on what roast color measures and how to read it, see Understanding Color Measurement in Coffee.

Link the result to your coffee

Assigning a coffee is optional. Tap Assign Beans in the Tool Controller and pick the roasted coffee in Bean Manager. The session then appears under Bean Insights on that coffee’s page, alongside any other CoffeeOS Tool sessions linked to it. Without it, the session stays only in the Roast Color Analyzer.

Roast Track curves are saved on OmniFlux. Swipe from right to left on the READY screen to open History.