DiFluid OmniFlux User Manual

May 2026, V1.0

Read this manual carefully so you can use this product correctly and safely.

  • Thank you for purchasing DiFluid OmniFlux Smart Roast Monitor Analyzer.
  • The information and specifications in this manual may change without notice as the product is improved.
  • Every effort was made to make this manual accurate at the time of writing. If you find an error, please let us know.

Video tutorial: watch how to set up and use OmniFlux

Using OmniFlux with CoffeeOS: this manual covers operating the device. To see your roast color in CoffeeOS, compare it with your roast profile and link results to your coffee, read the OmniFlux Complete Guide.

Every manual, as a web page and a PDF: DiFluid Product Manuals

Product Introduction

OmniFlux is built for color measurement. It combines several sensors with on-device analysis to monitor and record data through the whole roast.

It measures in two ways, real-time roast analysis and high-precision static color measurement, and adds tools that capture color, temperature, sound and other data for different situations.

Three Detection Modes

Color Test

  1. Quickly measures the average color value of the sample in front of it, for static color checks of roasted beans, ground coffee and similar samples.
  2. Takes several measurements at one point and gives their average, for daily calibration and comparison.

Roast Track

  1. Mounted next to the roaster, it records color, sound, temperature and other data in real time through the whole roast.
  2. Calibrates distance automatically within the working range and warns you if the distance is out of range.
  3. Works with PT100 temperature probes to record temperature curves alongside color, so it fits with traditional roast logging tools. (Temperature probes are sold separately.)

Cool Track

  1. Follows how the beans' color and temperature change while they cool after the roast, and plots time and temperature curves automatically.

Utility Tools

Besides the detection modes, OmniFlux has tools for use on site:

  1. Distance Meter: measures the distance between the device and the target in real time.
  2. Thermometer: shows the temperature where the red dot points, in real time, for example the room temperature.
  3. Temperature and Humidity Meter: records changes in the room's temperature and humidity.

Parts Overview

OmniFlux main unit, back, bottom and front

In the Box

OmniFlux package contents

Specifications

ModeDetectsItemSpecificationMethod and use
Color TestRoasted beans or ground coffee, level and evenly spreadColor value (beans and ground)Accuracy ±0.1
Range 0 to 150 Agtron
Static measurement with the Color Test Adapter; gives the average color value of the sample (beans and ground).
Roast TrackBeans inside the roaster drumColor value (beans)Accuracy ±1
Range 0 to 200 Agtron
Mounted outside the roaster window, records the color curve in real time during the roast. Temperature probes (sold separately) measure bean and drum temperature in real time.
Monitoring distance20 to 60 cm
AngleAny, with nothing blocking the optical path
Cool TrackBeans in the cooling trayInfrared temperatureAccuracy ±1 °C (0 to 320 °C)
Monitoring distance about 20 to 25 cm
Records the temperature curve of the beans while they cool.
ToolsNoneInfrared temperatureAccuracy ±1 °C (-70 to 380 °C)For roasting, quality control, equipment calibration, daily and outdoor use. Infrared laser ranging; the display shows the target spot's infrared temperature with room temperature and humidity.
Laser distanceRange 5 cm to 50 m
Accuracy ±3 cm
Battery and powerNoneBattery life6 hours or more
Charging time3 hours 30 minutes (5V/0.8A); can be used while charging
Discharge output5.0 to 5.2 V / 0.7 A

About the Power Supply

This product uses a lithium-ion battery. For safety, follow these precautions:

!If you will not use the device for a long time, charge it fully before storing it, and charge it again from time to time. Storing it with a low battery can damage the hardware.

!Do not take apart or modify the battery. It has built-in safety and protection devices; if they are damaged, the battery can overheat, smoke, burst or catch fire.

!Do not charge the battery near fire or in high heat, such as direct sunlight. Heat can trigger the protection device and stop charging. If the protection device is damaged, the battery can charge at abnormal current or voltage, which can cause overheating, smoke, bursting or fire.

!If you notice a strange smell, heat, discoloration, deformation or anything else unusual while using, charging or storing the battery, stop using the product at once. Continuing can lead to overheating, smoke, bursting or fire.

Power Usage Precautions

  1. Charge the device when it will not turn on or the battery is low.
  2. The charging port is on the bottom of the device. Use the included USB cable or any standard USB cable.
  3. No USB charging adapter is included. Use a standard USB adapter with a 5V/0.8A output.

Before You Start

Prepare the device as follows before you measure, so it works at its best and gives accurate data.

Battery Check

  1. Make sure the battery is charged. Start with more than 15%. The device can be used while charging.
  2. If the battery is low, charge it with the included cable and a stable power adapter before use.

Lens Cleaning

  1. Wipe the lens with a clean, soft, lint-free cloth. Fingerprints, dust and oil on the lens affect the data.

Caution

  • Do not look directly into the laser emitter. Laser light can damage your eyes.
  • Do not open the casing unless you are a qualified service technician. Doing so voids the warranty and may damage the device or stop it working.
  • DiFluid is not responsible for any damage caused by using this product (including accessories) incorrectly.
  • DiFluid is not liable for direct, indirect or other damages arising from defects in this product, even if advised of the possibility of such damages, nor for any third-party claims.
  • If anything about the device seems wrong, contact the store where you bought it.

Button Operation

OmniFlux has one physical button with several functions. A short press adapts to what the device is doing, so you can start or stop a measurement quickly.

When the Device Is Off

  1. The first time it is turned on after leaving the factory, press and hold for 3 seconds. A short press does not turn it on.
  2. Once a language is chosen at startup, either a short press or a 3-second press turns it on.
  3. After a factory reset, it goes back to step 1: only a 3-second press turns it on.

When the Device Is On

Press and hold

  • 3 seconds: the screen shows "Shut Down". Release the button to turn the device off.
  • 10 seconds: forces the device off, for when it does not shut down normally.

Short press

  • Outside a detection or calibration screen: returns to the main screen.
  • On a detection or calibration screen: the first press starts the current function (for example, a measurement); the next press stops it.

Touchscreen Operation

When OmniFlux is on, the touchscreen shows the READY screen.

READY screen
  1. App connection status: phone icon shown when connected to the app, crossed out when not connected.
  2. Battery level.
  3. Swipe from left to right to open the Menu.
  4. Swipe from right to left to open History.
  5. Tools (see Tools below).
  6. Tap anywhere on the screen to enter Color Test or Roast Track, chosen automatically from the distance.

How the Mode Is Chosen

Color Test starts at about 7 cm from the sample.

Starting Color Test mode

Roast Track starts at more than 10 cm.

Starting Roast Track mode

Settings

On the READY screen, swipe from left to right to open the Menu, then tap "Settings".

Settings screens
  1. The device's serial number (SN).
  2. Current firmware version. Connect to the CoffeeOS app regularly to check for new firmware and update the device: new firmware improves the experience, adds features and keeps performance stable.
  3. Scan this QR code to download CoffeeOS. If CoffeeOS is already installed, tap "Scan Device" at the top right of the app and scan the code to connect the device. CoffeeOS lets you track, save, record, manage, convert, calculate, analyze, study and explore your data. If the connection fails: on the device, go to Settings → Device Settings → Connection Settings and turn on "Bluetooth"; on your phone, turn on Bluetooth and location and allow every permission CoffeeOS asks for; then restart the device and CoffeeOS. If it still does not connect, contact the store where you bought the device.
  4. Connection settings: the Bluetooth switch must be on to connect to the app. With AirWave, the smoke eliminator, connected, OmniFlux remembers your settings and adjusts airflow through the roast (see Working with AirWave). Alert volume adjusts as needed. Turn this switch on when you connect the USB light.
  5. Screen brightness.
  6. Auto power-off time.
  7. Screen flip: turns the display 180°, for when the device has to be mounted upside down.
  8. Date and time format.
  9. Language: English, Japanese or Simplified Chinese.
  10. Factory reset: press and hold until the progress bar fills to clear all settings and history.
  11. Roast Track settings (see Live Roast Tracking).
  12. Color standard: SCA or COMMON.
  13. Temperature unit: °C or °F.

First Use

Before using OmniFlux for the first time, do these two things, in this order:

  1. Color Calibration
  2. New Roaster Profile

After that, the everyday sequence is: load the beans (Bean Placement), Color Test, Live Roast Tracking, and calibration of the roast result when needed.

Color Calibration

Color calibration keeps OmniFlux's color measurement accurate.

Calibration

  • Calibrate before you use OmniFlux for the first time.
  • Calibrate once a week after that.
1. Take the two-color calibration card out of the box and lay it on the table. Do not scratch the card's detection areas.
Two-color calibration card
2. On the READY screen, swipe left to open the menu. Tap "Calibration" → "Color Test Calibration".
3. Enter the two reference values printed on the card and tap "Next".
Entering the reference values
4. Fit the Color Test Adapter under the device's optical module.
Fitting the Color Test Adapter
5. Place the device on reference value ① on the card and tap "Measure ①".
Measuring reference value 1
6. Move OmniFlux to reference value ② on the card and tap "Measure ②". When the check mark lights up, tap "Calibrate" to finish.

Stand Installation

1. Take out the main bracket and fix it to the edge of the table as shown. Tighten it securely.
Fixing the bracket to the table
2. Take out the connector. Remove the nut, then fit the two parts of the connector to the main bracket: take out the nut, place it on the main bracket, then press the connector firmly into the top of the bracket.
Fitting the connector
3. Twist the two points shown by the arrows until tight.
Tightening the connector
4. Take out the quick-release head, line up its locating post with the hole on the bottom of the device, and tighten it with the screwdriver provided.
Fitting the quick-release head
5. Hold down the button on the quick-release head and slide the device onto the bracket. Installation is complete.
Attaching the device
6. To remove the device, hold down the button on the quick-release head and slide it out. Do not knock the bracket, or it may move.
Removing the device

New Roaster Profile

  • Each roaster usually needs to be set up only once.
  • You can create several profiles for the same roaster if needed, for example for different roast levels.
  • A different roaster needs a new profile.

Before you start

  • Do not move the bracket once the profile is set up.
  • During roast tracking, always keep the same angle and distance as when the profile was set up.
1. On the READY screen, swipe left to open the menu. Tap "Calibrate" → "Roast Track Calibration".
2. Tap "New Roaster Profile" and enter the roaster's name.

i"Default" is the factory profile. You can choose "Default" from the drop-down on the Roast Track screen and start roasting straight away, then save the result under your roaster's name.

3. Set the bracket's height and angle, then check that all three knobs on the bracket are tight. A loose knob lets the position shift. (See Stand Installation.)
4. Fit the device on the bracket.
5. Aim the red laser dot level or slightly downward at the center of the roaster window, where the beans are densest, so it reads the largest area of beans.
6. Start the roaster and heat it as usual. The drum must finish preheating first.
7. Charge the green beans. They must cover the window completely. Tap "Next".
8. When the starting color value is steady, tap "Record Green Beans". The device needs nothing more until the roast ends.
Calibration process screen
9. At the end of the roast, tap "Record Roasted Beans" and drop the beans immediately.

!If you do not drop the beans right after tapping "Record Roasted Beans", the device may record the wrong final color.

10. Take the device off the bracket without changing the bracket's angle (see Stand Installation).
11. Tap the "Roasted Bean Color Value" field, fit the Color Test Adapter to OmniFlux, and measure the cooled beans. Enter the final color value. For an accurate reading, follow Bean Placement. The roaster profile is complete.

Bean Placement

iIf you don't need the highest accuracy, use the standard method: level the beans with the scraper. There is no need to pick out beans or turn them silver skin down.

iFor the most accurate measurement, spread the beans this way: take out defective beans and beans clearly different in color, and level the beans with the top of the sample tray.

iIf the silver skin is clearly a different color, turn the beans over by hand so the silver skin faces down.

iSpread and measure 2 or 3 times. When the color values agree, record the result.

Color Test

Before you measure

  • Make sure the lens is free of fingerprints and dust. Wipe it with a soft, lint-free cloth.
  • Calibrate from the menu before measuring, ideally once a week (see Color Calibration).
  • Fill the sample tray and level the sample.
  • Samples: roasted beans or ground coffee, each in its matching tray.
  • This mode measures a still sample and needs the Color Test Adapter (included).

iFor each new batch, tap "New Test" first.

i"Add Test" adds each new result to the ones before it and shows the average.

SituationTapExample
You changed the sample (another batch of beans or grounds)New TestMeasure the first batch and tap "New Test": the result is 80.
You finished a full round and are starting a new batchNew TestMeasure the second batch and tap "New Test": the result is 130.
You are measuring one batch in parts (for example 2 g at a time, 5 times, 10 g in all) and want the averageAdd TestMeasure the third batch and tap "Add Test": it shows the average of the second and third batches, not the first.
You want the average of one sample measured several ways, such as turning the tray or spreading the grounds againAdd TestMeasure the fourth batch and tap "Add Test": it shows the average of the second, third and fourth. For a new batch, tap "New Test" and earlier results are no longer included.

Steps

1. Fit the Color Test Adapter under the device's optical module.
Fitting the Color Test Adapter
2. Fill the tray with the sample and level it with the scraper: the shallow groove for grounds, the deep groove for beans. See Bean Placement for the most accurate results.
Filling and leveling the sample
3. Place the device with the adapter fitted onto the sample tray.
Placing the device on the tray
4. Line it up with the edge of the metal tray. For a new batch, tap "New Test" first.
Starting a new test
5. When the result shows, tap "Add Test" to add the next measurement. The screen shows the average of this result and the results since the last "New Test".
Adding a test

Live Roast Tracking

The sequence: set up before the roast → during the roast → before the drop → cooling → confirm color deviation → adjust the drop time (curve trimming).

  • Keep the measuring distance and angle the same as when the roaster profile was set up.
  • Measure bean color only after the beans have cooled (see Bean Placement and Color Test).
  • If roast results drift, calibrate promptly (see Confirm Color Deviation below).
  • Clean the roaster window regularly so oil buildup does not affect the readings.

Before the Roast

1. Mount the device on the stand at the same angle and distance as when the roaster profile was set up.

iIf your roaster has a small window, keep the stand fixed in place, so it is not moved and the angle does not have to be found again.

2. Set the parameters: on the READY screen tap anywhere to open Roast Track, then pull down from the top to open the pull-down settings.
Opening Roast Track and the pull-down settings
Pull-down settings
  1. Roaster Profile: choose the profile for the roaster you are about to use.
  2. Auto Start / Stop: On: tracking starts by itself when you charge the beans and continues past the target color; when the next batch is charged, the last result is saved and a new tracking starts. Off: tap the start icon at the bottom right after charging to begin tracking; tracking stops when the target color is reached.
  3. Target Color Alert: recommended. An alert sounds when the target color is reached.
  4. Temperature probe: turn this on yourself if a probe (sold separately) is connected.
3. Set the target color by pressing and holding the + and - buttons. When the beans reach it, the device tells you to drop.
Setting the target color
4. Start the roaster and heat it as usual. The drum must finish preheating first.
5. Tap the start icon at the bottom right and charge the beans.

iOr turn on "Auto Start / Stop" in the pull-down settings so tracking starts without tapping.

6. Roast tracking begins.

During the Roast

  • Yellowing, first crack and second crack are marked automatically, and you can also mark them yourself.
Roast milestones on the curve
  • You can change the target color at any time.
Changing the target color during the roast

Before the Drop

  • Watch the color trend to judge the drop in advance and reduce deviation.
  • Note the time or color value at the moment you drop, so you can calibrate accurately afterwards.
  • When the alert sounds, drop the beans straight away. You don't need to touch the device; just work the roaster. After dropping, tap "End Roasting" in the pop-up.

Cooling

1. After the drop, tap the snowflake icon at the bottom right to track the beans' temperature with the infrared sensor while they cool.

Confirm Color Deviation (Calibration)

iThis step corrects color deviation in tracking. With correct setup and even beans, deviation stays within ±1, although beans of uneven color can still cause larger deviation.

Check that the time or color value on the left of the results page matches what you noted at the drop. If it matches, carry on. If not, trim the curve first (see Curve Trimming below), then calibrate.

Results page
1. Put the cooled beans in the sample tray (see Bean Placement).
2. Fit the Color Test Adapter to the device and place it over the tray to measure the color (see Color Test).
3. Confirm the real color value and correct any deviation. Two examples show how it works:
Recalibrating a roaster profile

Example 1. Tap ① drop color (say 75) → ② roaster name → ③ Recalibrate → ④ Final roast color → ⑤ Measure (result 78) → ⑥ Save. The deviation was 3. Value ① now reads 78, and the next roast shows a more accurate roast color.

Example 2. Tap ① drop color (say 75) → ② roaster name → ③ Recalibrate → ④ Final roast color → ⑤ Measure (result 75) → ⑥ Save. The deviation was 0. Value ① stays at 75.

Curve Trimming

Tap the results page to open the trimming and calibration screen.

Curve trimming screen

Any one of methods ①, ② and ③ trims the curve:

  1. Drag the dashed line to the drop time or color you noted, then tap ⑥ to cut off the curve after the drop.
  2. Enter the drop color you noted to find the drop point, then tap ⑥ to cut off the curve after it.
  3. Enter the drop time you noted to find the drop point, then tap ⑥ to cut off the curve after it.
  4. Zoom in or out on the curve around the dashed line.
  5. Press and hold to move the time left or right.

Working with AirWave

Where to connect: READY screen → swipe right → Settings → Device Settings → Connection Settings → AirWave Connection.

Connecting

1. Turn on both AirWave and OmniFlux.
2. On AirWave's home screen, press and hold the knob to open settings.
3. Turn the knob right to "Mode", then press briefly to choose "Auto Mode".
4. Keep turning right to "Wireless Connection", then press briefly to turn it "On".
5. OmniFlux shows the smoke eliminator's name. Tap it to connect, then swipe right on the screen to leave settings.
6. Press and hold AirWave's knob to leave settings and return to the home screen, ready to roast.

The First Roast Together

1. Open the Roast Track screen on OmniFlux.
2. While you roast, set AirWave's fan speed by hand to suit the smoke, for example at yellowing and first crack.
3. Each change you make is recorded automatically.

Every Roast After That

1. AirWave applies the recorded fan speeds automatically.
2. You can still turn the knob to change the fan speed during the roast. Your latest setting is used from then on.

Tools

On the home screen, tap the tool icon at the top right to open the tools.

Tool Mode screen
ToolWhat it doesUses
①Distance meterMeasures distance with the infrared laser. Range 5 cm to 50 m, accuracy ±3 cm.Roasting: measures the distance between the roaster and the sampling point, to help with calibration. Home: measuring rooms, interior design, hanging pictures.
②ThermometerRange -70 to 380 °C. Shows the infrared temperature of one spot in real time.Roasting: quick surface temperatures of equipment and the cooling tray. Home: cooking pots, air conditioning vents.
③Temperature and humidity meterShows room temperature, humidity, air pressure and altitude.Roasting: records the roastery's temperature and humidity and the roasting conditions. Home: room conditions, wine cellar and storage.

Roast Color Reference

How Agtron values map to the two color standards OmniFlux can show (set in Settings).

AgtronCOMMONSCA
0 to 30EspressoVery Dark
above 30 to 40FrenchDark
above 40 to 50Full CityMedium Dark
above 50 to 60CityMedium
above 60 to 70DarkMedium Light
above 70 to 80MediumLight
above 80 to 90CinnamonVery Light
above 90 to 150LightExtremely Light

Cleaning and Maintenance

Clean the lens every day before measuring. Wipe it with a clean, soft, lint-free cloth so fingerprints, dust and grease do not affect the data.

Warranty Policy

Faults that occur under normal use are repaired free of charge by DiFluid for one year from the date of purchase.

Not Covered by Warranty

  1. Products bought outside DiFluid's official authorized sales channels, or without a warranty card and valid invoice.
  2. Products past the warranty period.
  3. Faults or damage caused by misuse, modification, disassembly, repair or flashing firmware yourself.
  4. Faults or damage caused by shipping, moving or dropping the product after purchase.
  5. Accidental or user damage, such as liquid getting in, the wrong input voltage, crushing, or scratches on the screen or casing.
  6. Faults or damage caused by force majeure, such as fire, earthquake or flood.

Compliance Information

Federal Communications Commission (FCC) Radiation Exposure Statement

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.

NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and receiver.
  • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.

When using the product, maintain a distance of 5 mm from the body to ensure compliance with RF exposure requirements.

CE Compliance Notice

  1. Due to the enclosure material used, the product shall only be connected to a USB interface of version 2.0 or higher. Connection to so-called power USB is prohibited.
  2. Do not use the device at too high or too low a temperature, and never expose it to strong sunlight or a very wet environment. The suitable temperature for the product and accessories is 0 to 45 °C.
  3. Warning: replacing a battery with an incorrect type can defeat a safeguard; disposing of a battery into fire or a hot oven, or mechanically crushing or cutting a battery, can result in an explosion; leaving a battery in an extremely hot environment can result in an explosion or the leakage of flammable liquid or gas; a battery subjected to extremely low air pressure may result in an explosion or the leakage of flammable liquid or gas.
  4. This device is a low-power radio transmitter and receiver. As recommended by international guidelines, the device meets applicable national SAR limits of 2.0 W/kg (10 g). 10 g SAR: meets the low-power exclusion level; a SAR test is not required.
  5. Operating frequency range: Bluetooth 2402 MHz to 2480 MHz, 9.48 dBm. This product can be used across EU member states.

EU Regulatory Conformance. Hereby, Shenzhen Digitizing Fluid Technology Co., Ltd. declares that this device is in compliance with the essential requirements and other relevant provisions of Directive 2014/53/EU.

Shenzhen Digitizing Fluid Technology Co., Ltd.
Room 1602, Jinhua Building, Longhua District, Shenzhen, China
(+86) 0755-23761557