DiFluid R1 Extract User Manual
Pocket Refractometer, Quick Startup Guide
Read this manual carefully so you can use this product correctly and safely.
- Thank you for purchasing DiFluid R1 Extract Pocket Refractometer.
- The information and specifications in this manual may change without notice as the product is improved.
- Every effort was made to make this manual accurate at the time of writing. If you find an error, please let us know.
Using R1 Extract with CoffeeOS: this manual covers operating the device. To use your TDS readings in CoffeeOS, read the R1 Extract Complete Guide.
Every manual, as a web page and a PDF: DiFluid Product Manuals
Overview

| Part | What it does |
|---|---|
| Sampling tank | Holds the liquid sample over the prism at its base. |
| Shell | The device housing. |
| Button | Turns the device on, starts a measurement and starts calibration with clean water. |
| Status | Status icons on the display. |
| TDS / Concentration | The TDS (concentration) reading, in %. |
| Temperature | The sample temperature, in °C. |
| Refractive index | The refractive index of the sample (N). |
| USB-C charging port | Charges the device. |
Accessories

- Type-C charging cable
- Cleaning cloth
- Drawstring bag
- Droppers
Specifications
As printed on the label on the back of the device:
| Item | Specification |
|---|---|
| Name | DiFluid Coffee |
| Model | DFT-F10V55H23 |
| TDS range | 0 to 26% |
| TDS precision | ±0.03 |
| Input | 5 V, 350 mA |
| Sample temperature | 5 to 45 °C (see Notes) |
Calibration
Calibration sets the zero point with clean water.

- Press the button to turn the device on. The startup screen appears.
- Drip the water you brew coffee with onto the sampling tank until it completely covers the prism at the bottom.

- Press the button once briefly, then press and hold it until the screen shows the calibrating message. The device is now calibrating.
- Calibration is complete when the screen shows a TDS of 0.0%.
Measurement

- Press the button briefly to turn the device on.
- Drip the liquid you want to measure, such as coffee, onto the sampling tank until it completely covers the prism at the bottom.

- Press the button briefly to measure.
- The result appears on the screen.
If the liquid is hot and the reading does not settle, measure several times.
Cleaning the Sampling Tank
Step 1

Rinse the prism at the bottom of the sampling tank with clean water.
Step 2

Dry the sampling tank with the cleaning cloth or a paper towel.
App Connection
Step 1

Open DiFluid app and select Scan code to connect device.
Step 2

Scan the QR code on the back of the device to connect it to the app.
Notes
- Clean and wipe the sampling tank before and after every measurement.
- Measure liquids between 5 and 45 °C. Outside this temperature range, results will be inaccurate.
- Drip enough liquid to completely cover the prism.
- If results are unstable or deviate too much, rinse the prism with clean water, wipe it clean and calibrate the device again.
- Before charging, make sure the device, including the charging port, is completely dry.
Compliance Information
FCC Compliance Notice
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
- This device may not cause harmful interference.
- This device must accept any interference received, including interference that may cause undesired operation.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
NOTE: The manufacturer is not responsible for any radio or TV interference caused by unauthorized modifications to this equipment. Such modifications could void the user's authority to operate the equipment.
RF Exposure
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with a minimum distance of 20 cm between the radiator and your body. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.
EU Compliance Statement
Hereby, Shenzhen Digitizing Fluid Technology Co., Ltd. declares that the radio equipment type DFT-F10V55H23 is in compliance with Directive 2014/53/EU. The full text of the EU declaration of conformity is available at the following internet address: https://www.digitizefluid.com/support/document
RF Exposure Information
The EIRP power of the device at maximal case is below the exempt condition, 20 mW, specified in EN 62479:2010. RF exposure assessment has been performed to prove that this unit will not generate harmful EM emission above the reference level as specified in EC Council Recommendation (1999/519/EC).
Shenzhen Digitizing Fluid Technology Co., Ltd.
Room 1602, Jinhua Building, Longhua District, Shenzhen, China
(+86) 0755-23761557